SLAN-96P Real-Time PCR System

Specifications
Product model: 	SLAN-96P
Sample capacity:	96 wells (dual 48/48 reaction module)
Sample size:       	15-50mL
Consumable:  	        0.2mL PCR tubes, 8-tube strips, 48-well plates
Parallel running mode:	Dual reaction blocks/running 2 tests independently
Temperature range: 	4-99°C
Max ramp rate:          	4.0°C/sec
Temperature accuracy:	        ±0.1°C
Temperature uniformity:	±0.1°C
Thermal cycling system: 	Peltier-based system
Temperature control mode:	Tube control/block control
Light Source:   	     LED (maintenance-free)
Detector:  	             High sensitivity photoelectric sensor
Sensitivity: 	             1 copy
Linearity range:       100-1010
Resolution:	Can discriminate between 1000 copies and 2000 copies
Excitation:	CH1 470nm    CH2 530nm    CH3 580nm
CH4 630nm   CH5/CH6   custom-made
Emission:	CH1 510nm    CH2 565nm    CH3 620nm
CH4 665nm   CH5/CH6   custom-made
Dyes and probes:	CH1 FAMTM /SYBRGreen®   CH2 VIC®/HEX/J0ETM /TET
CH3 ROX/TexasRed®   CH4 CY5TM
Hot-lid	30°C~108°C (default105°C, adjustable) electronic automatic hot-lid
Power supply:            	230VAC, 50Hz
Power consumption:	850VA
Dimensions:               	380mm x 520mm x 250mm (WxDxH)
Weight: 	       18Kg
Communication:	RS232, USB
Qualification:	CE
Precise Temperature Control System:

1. The SLAN system adopts Peltier technology which can accurately and quickly control increases and decreases of temperature with a precision of ±0.1℃ and features long service life, no noise and no pollution, etc.;

2. Multiple-point temperature control technology ensures the uniformity between different wells. The temperature correction technology guarantees the accuracy of temperature and thus ensures the accuracy of assay result;

3. PCR tube temperature-control technology guarantees the accuracy of actual temperature of reagent in the tube and enables the SLAN to easily adjust for different reactions using different volumes.
Sensitive Photoelectric Detection System

1. High sensitivity photoelectric detector with no background noise which is fast and stable;

2. Lifetime maintenance-free ultra-bright LED cold light source with a large signal value and high stability;

3. Unique optical fiber conduction technology that greatly improves fluorescence collection efficiency and avoids fluorescence interference between neighboring tubes;

4. The system automatically selects best gain value for each tube without the need for user configuration. The SLAN®has an extremely broad measuring range for fluorescence;

5. Extremely high detection sensitivity with an extremely low fluorescence background value. It’s not necessary for users to make optical correction or background correction.

SLAN-96P Multi-tasking Software Interface (analysis screen)

Unique Electronic Automatic Hot Lid System

1. Flexible heating technology ensures uniform hot lid temperature;

2. The electronic automatic hot lid has an adjustment function for different 0.2ml reaction tubes to ensure even tube pressure;

3. Once the assay starts, the hot lid has automatic locking function to prevent assay failure following accidental opening.